• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. PRECOAT FEMORAL DRILL W/STOP; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. PRECOAT FEMORAL DRILL W/STOP; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Model Number 00584301800
Device Problem Defective Device (2588)
Patient Problem No Patient Involvement (2645)
Event Date 06/02/2020
Event Type  malfunction  
Event Description
It was reported that upon inspection the drills were found dull.No injury, no delay, noticed after case.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.The stated failure was confirmed through a visual inspection that the femoral drill is dull and has burrs from contact with mating device.This device shows significant signs of wear/usage.This device was manufactured in 2016.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECOAT FEMORAL DRILL W/STOP
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10192373
MDR Text Key196379540
Report Number1020279-2020-02706
Device Sequence Number1
Product Code HWE
UDI-Device Identifier00885556621318
UDI-Public00885556621318
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 12/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number00584301800
Device Catalogue Number00584301800
Device Lot Number63279536
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2020
Initial Date Manufacturer Received 06/02/2020
Initial Date FDA Received06/24/2020
Supplement Dates Manufacturer Received11/17/2020
Supplement Dates FDA Received12/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-