MEDTRONIC, INC. ACHIEVE ADVANCE MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
Model Number 2ACH20 |
Device Problems
Entrapment of Device (1212); Material Integrity Problem (2978)
|
Patient Problem
Physical Entrapment (2327)
|
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during a cryo ablation procedure, the mapping catheter was rotated deep inside of a pulmonary vein.The catheter cinched up on itself, and started to grab into parenchyma tissue.The catheter was able to be untangled, and was successfully utilize for the remainder of the case.After the case, upon removal, no damage was noted on the catheter.The case was completed.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|