• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 SL FOLEY SWIVEL SILICONE TRICOT 25BX; STATLOCK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 SL FOLEY SWIVEL SILICONE TRICOT 25BX; STATLOCK Back to Search Results
Catalog Number FOL0102CE
Device Problems Disconnection (1171); Detachment of Device or Device Component (2907)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 05/20/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the rotating clamp of the statlock came off from the mount easily.
 
Manufacturer Narrative
The reported event was confirmed, however the cause was unknown.Visual inspection noted that one opened and 8 unopened statlock devices were received.Visual evaluation noted that the one opened sample had its base completely disconnected from the adhesive pad.The unopened samples had their bases adhered properly to the adhesive pad.Although the reported event was confirmed, the root cause could not be determined.A potential root cause for this failure mode could be due to worn out brush.The product was not used for the diagnosis or treatment.The product had failed to meet the specifications, and was influenced by the reported failure.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "statlock® foley stabilization device.Read carefully before use.Safety and efficacy considerations: the statlock® device is for single use only.Sterilized using ethylene oxide.Sterile unless package is opened or damaged, except for any individually packaged components within the pouch which are not labeled as sterile.These components are not terminally sterilized.Do not alter the statlock® device or components.Procedure must be performed by trained personnel with knowledge of anatomical landmarks, safe technique and potential complications.Not made with natural rubber latex.Indications for use: the statlock® device is a stabilization device for compatible catheters.Contraindications: known tape or adhesive allergies.Warnings and precautions: 1.Do not use the statlock® device where loss of adherence could occur, such as with a confused patient, unattended access device, diaphoretic or non-adherent skin.2.Observe universal blood and body fluid precautions and infection control procedures, during application and removal of the statlock® device.3.Minimize catheter manipulation during application and removal of the statlock® device.4.Daily maintenance.A.The statlock® device should be assessed daily and changed when clinically indicated, at least every seven days.B.If pad becomes soiled, wash with soap / water, saline or hydrogen peroxide.Do not use alcohol or prepackaged bathing systems, which could lead to early lifting.C.If showering / bathing, cover with plastic wrap or waterproof dressing.D.Conduct skin assessment prior to application and repeat daily per facility protocol.E.Use clinical judgment on the removal of the statlock® stabilization device if the patient experiences any fluid shifts that may interfere with skin integrity.Application technique: prep: 1.Place foley catheter into retainer.Directional arrow should point towards catheter tip, and balloon inflation arm should be next to the hinge.2.Close lid, being careful to avoid pinching the catheter.3.Identify securement site by laying the device retainer on the front of the thigh, leaving 2.5 cm of catheter slack between insertion site and the statlock® device retainer.4.After placing the statlock® stabilization device off to the side, cleanse and degrease the securement site with alcohol per hospital policy.Let skin dry.5.Apply skin protectant, in direction of hair growth, to area larger than securement site.Allow to dry completely (10-15 seconds).6.Using permanent marker, write initials and date of application on the statlock® device anchor pad.Note: always secure catheter into the statlock® device retainer before applying adhesive pad on skin.Place and peel: 7.Align the statlock® stabilization device over securement site leaving 2.5 cm of catheter slack.Make sure leg is fully extended.8.While holding the retainer to keep the pad in place, peel away paper backing, one side at a time and place tension-free on skin.Removal technique: disengage: 1.Open retainer by pressing release button with thumb, then lift to open.2.Remove foley catheter from the statlock® device.Dissolve: 3.Wipe the edge of the pad using at least 5-6 alcohol pads until a corner lifts.Then continue to stroke undersurface of pad with alcohol to dissolve adhesive pad away from skin.Do not pull or force pad to remove." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that the rotating clamp of the statlock came off from the mount easily.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SL FOLEY SWIVEL SILICONE TRICOT 25BX
Type of Device
STATLOCK
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key10200594
MDR Text Key196917979
Report Number1018233-2020-04125
Device Sequence Number1
Product Code EYJ
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Type of Report Initial,Followup
Report Date 08/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/28/2022
Device Catalogue NumberFOL0102CE
Device Lot NumberJUDSF211
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/23/2020
Initial Date Manufacturer Received 06/05/2020
Initial Date FDA Received06/26/2020
Supplement Dates Manufacturer Received08/03/2020
Supplement Dates FDA Received08/18/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-