• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. OES CYSTONEPHROFIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. OES CYSTONEPHROFIBERSCOPE Back to Search Results
Model Number CYF-5
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/21/2020
Event Type  malfunction  
Manufacturer Narrative
The subject device has not been returned to omsc but was returned to olympus (b)(4).(b)(4) sent the subject device to a third party laboratory for microbiological testing.As a result of the testing, the sample collected from all channels of the device tested positive for unspecified microbes (<1ufc/100ml).The testing result cleared the french guideline.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that as a result of routine microbiological testing by the user facility, the sample collected from all channels of the subject device tested positive for staphylococci coagulase (2ufc/100ml).The device had been reprocessed using peracetic acid.There was no report of infection associated with this report.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Olympus medical systems corp.(omsc) reviewed the manufacturing history (dhr) of the device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES CYSTONEPHROFIBERSCOPE
Type of Device
CYSTONEPHROFIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10204777
MDR Text Key199833310
Report Number8010047-2020-03705
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
PMA/PMN Number
K032092
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 09/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-5
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/02/2020
Initial Date FDA Received06/29/2020
Supplement Dates Manufacturer Received08/24/2020
Supplement Dates FDA Received09/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-