Device was received for evaluation.Evaluation determined that the device was found to have foreign object on the channel.Biopsy was found leaking and the ¿a-rubber¿ glue was found worn out.Objective lens were found pitted and scratches were found on the edge of the device.Light guide was found squashed.Plastic around the biopsy opening was determined with scratches and burnt/melted.The identified parts were replaced, device was repaired.Once completed, the device was tested and passed all required testing and specifications.Based on evaluation findings, the reported issue was confirmed.The likely cause of the issue was attributed to maintenance and or mishandling.
|
This supplemental report is being submitted to inform that upon further review this is not a reportable malfunction.Per the legal manufacturer there is no potential for this issue to cause or contribute to death or serious injury if the malfunction were to recur.Please see updated sections: g4, g7, h2, h3, and h10.
|