Model Number PM3210
Device Problems Failure to Interrogate (1332); Wireless Communication Problem (3283)
Patient Problems No Known Impact Or Consequence To Patient (2692); Insufficient Information (4580)
Event Date 06/17/2020
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that patient¿s device was not able to be interrogated.The interrogation was attempted using inductive telemetry, different programmer, clearing the environment to remove effect of electromagnetic interference (emi), magnet, radio frequency telemetry test.All troubleshooting failed.The decision was made to interrogate the device in a different room.If interrogation fails, the device replacement was discussed.Further information was requested but was not received.
 
Event Description
New information received notes that the device was explanted and replaced.Elective replacement indicator (eri) was also noted.
 
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Manufacturer Narrative
The reported field event of failure to interrogate was not confirmed in the laboratory.The device was tested on the bench, and no anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHEM RF CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
MDR Report Key10211387
MDR Text Key196923655
Report Number2938836-2020-07174
Device Sequence Number1
Product Code NKE
UDI-Device Identifier05414734503785
UDI-Public05414734503785
Combination Product (y/n)N
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup,Followup
Report Date 02/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2014
Device Model NumberPM3210
Device Catalogue NumberPM3210
Device Lot Number3782427
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/17/2020
Initial Date FDA Received06/30/2020
Supplement Dates Manufacturer Received11/12/2020
02/18/2021
Supplement Dates FDA Received12/05/2020
02/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-