• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG; STANDARD CART W/ ISOLATION TRANSFORMER; CART, EMERGENCY, CARDIOPULMONARY (EXCLUDING EQUIPMENT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG; STANDARD CART W/ ISOLATION TRANSFORMER; CART, EMERGENCY, CARDIOPULMONARY (EXCLUDING EQUIPMENT) Back to Search Results
Catalog Number 0240099011
Device Problems Sparking (2595); Electrical Shorting (2926)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/01/2019
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event Description
It was reported that there was sparking on plugging in the device.
 
Event Description
It was reported that there was sparking on plugging in the device.
 
Manufacturer Narrative
The product was not returned for investigation and the reported failure mode was not confirmed.The failure mode will be monitored for future reoccurrence.Alleged failure: sparking probable root cause: too much heat transformer design cart design manufacturing nonconformity (tekna) use error the device manufacture date is not known.81.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG; STANDARD CART W/ ISOLATION TRANSFORMER
Type of Device
CART, EMERGENCY, CARDIOPULMONARY (EXCLUDING EQUIPMENT)
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key10213549
MDR Text Key197297234
Report Number0002936485-2020-00250
Device Sequence Number1
Product Code BZN
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 09/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0240099011
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/02/2020
Initial Date FDA Received06/30/2020
Supplement Dates Manufacturer Received06/02/2020
Supplement Dates FDA Received09/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-