MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII P/S CONS HOUSING REAM DOME; PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII P/S CONS HOUSING REAM DOME; PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY Back to Search Results
Catalog Number 71440145
Device Problem Dull, Blunt (2407)
Patient Problem No Patient Involvement (2645)
Event Date 06/08/2020
Event Type  malfunction  
Event Description
It was reported that the genesis ii articular insert spacer trial 1-2 13 mm, the genesis ii p/s cons housing reamer dome and the femoral implant impactor are worn and torn from use.No patient was involved.No other complications were reported.
 
Manufacturer Narrative
The device, intended use in treatment, was returned for evaluation.A visual inspection confirmed the gii p/s cons housing ream dome is dull.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII P/S CONS HOUSING REAM DOME
Type of Device
PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10213625
MDR Text Key197082273
Report Number1020279-2020-02843
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 08/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71440145
Initial Date Manufacturer Received 06/08/2020
Initial Date FDA Received06/30/2020
Supplement Dates Manufacturer Received07/30/2020
Supplement Dates FDA Received08/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-