SMITH & NEPHEW, INC. GII P/S CONS HOUSING REAM DOME; PROSTH, KNEE, PATE/FEMOROTIBIAL, SEMI-CONST, UNCEM, PORO, COAT, POLY/METAL/POLY
|
Back to Search Results |
|
Catalog Number 71440145 |
Device Problem
Dull, Blunt (2407)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 06/08/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that the genesis ii articular insert spacer trial 1-2 13 mm, the genesis ii p/s cons housing reamer dome and the femoral implant impactor are worn and torn from use.No patient was involved.No other complications were reported.
|
|
Manufacturer Narrative
|
The device, intended use in treatment, was returned for evaluation.A visual inspection confirmed the gii p/s cons housing ream dome is dull.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.
|
|