• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION BLOOD RECIPIENT SET; SET, BLOOD TRANSFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION BLOOD RECIPIENT SET; SET, BLOOD TRANSFUSION Back to Search Results
Catalog Number JC7751
Device Problem Disconnection (1171)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the tubing of y-type blood/soln set separated from the luer.This issue was noted during blood transfusion.The set was replaced.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information was added to h3, h4 and h6 (replace 4115 with 10).H10: the actual device was discarded; however, a photograph of the sample was provided for evaluation.Visual inspection was performed to the photograph using the naked eye which observed that the tubing was separated from the luer lock component.Due to the nature of sample, no functional testing could be performed.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLOOD RECIPIENT SET
Type of Device
SET, BLOOD TRANSFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10214361
MDR Text Key197170411
Report Number1416980-2020-03661
Device Sequence Number1
Product Code BRZ
UDI-Device Identifier00085412152738
UDI-Public(01)00085412152738
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberJC7751
Device Lot NumberDR19H10019
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/04/2020
Initial Date FDA Received06/30/2020
Supplement Dates Manufacturer Received07/10/2020
Supplement Dates FDA Received08/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-