• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRN II BCS LCK FEM IMPLANT IMPACT; PROSTHESIS, TOE, CONSTRAINED, POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRN II BCS LCK FEM IMPLANT IMPACT; PROSTHESIS, TOE, CONSTRAINED, POLYMER Back to Search Results
Model Number 74012812
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 06/08/2020
Event Type  malfunction  
Event Description
It was reported that upon inspection the instrument was found in need of replacement due to wear and tear of use.The plastic tip has developed a chip.No case involved.
 
Manufacturer Narrative
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The cover on the cam arm is fractured, rendering the device inoperable.The device was manufactured in 2016 and shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRN II BCS LCK FEM IMPLANT IMPACT
Type of Device
PROSTHESIS, TOE, CONSTRAINED, POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10216149
MDR Text Key197745473
Report Number1020279-2020-02853
Device Sequence Number1
Product Code KWH
UDI-Device Identifier00885556003541
UDI-Public00885556003541
Combination Product (y/n)N
PMA/PMN Number
K111711
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 01/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number74012812
Device Catalogue Number74012812
Device Lot Number16ESM0025
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2020
Initial Date Manufacturer Received 06/08/2020
Initial Date FDA Received06/30/2020
Supplement Dates Manufacturer Received06/08/2020
Supplement Dates FDA Received01/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-