The patient's charging device would not connect to the ipg (neurostimulator).Several different charging devices were used in an attempt to connect to the ipg.Also attempted to rotate the charging device in different orientations.The issue was unable to be resolved.The implanting physician mentioned during the post-operation follow up appointment that they implanted the ipg no more than 2 cm.The physician palpated for the ipg and was able to locate it.Impedances were checked and observed to be fine.The physician decided to schedule the patient for a pocket revision on (b)(6) 2020.The clinical specialist was able to meet with the patient prior to the surgery on (b)(6) 2020.The patient was finally able to get a connection tone between the charging device and ipg.On (b)(6) 2020, the physician went ahead with the surgery to replaced the patient's ipg with a new ipg.
|