The patient was undergoing a thrombectomy procedure in the left anterior descending artery (lad) using a penumbra system aspiration pump max 110 (pump max).During the procedure, a hospital staff accidentally connected the aspiration tubing (tubing) directly to the pump max rather than the canister.Consequently, blood was aspirated into the pump max; therefore, the pump max was disconnected.The procedure was completed using balloon angioplasty and stent.There was no report of an adverse effect to the patient.
|
Results: blood was present on the pump housing, and therefore, no functional testing was performed.Conclusions: evaluation of the returned pump max confirmed blood was aspirated inside the vacuum pump.The complaint reported that the aspiration tubing was connected directly to the pump max vacuum inlet, rather than the canister, supplied by penumbra.If fluid is aspirated into the vacuum assembly, the pump may not function properly.Penumbra pumps are visually inspected and functionally tested during incoming quality inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
|