• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. UNKN IV3000; TAPE AND BANDAGE, ADHESIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. UNKN IV3000; TAPE AND BANDAGE, ADHESIVE Back to Search Results
Catalog Number UNKN05003901
Device Problem Device Slipped (1584)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/01/2013
Event Type  malfunction  
Event Description
**literature case** in a scientific publication: "clinical observation of needle-free closed joint for central venous catheter infusion in neurosurgery".It was reported that 246 patients were using iv3000 (smith&nephew) for fixation and had problems due to infusion set slippage.There is no information regarding to the outcome of the patients or the treatment provided for this complication.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
We have now concluded our investigation for the complaint received.A review of the batch manufacturing records could not be performed as no part or lot numbers were provided, however, there are no indications to suggest that the device did not meet specifications upon release into distribution.A complaint history review was carried out across all iv3000 products, there have been further complaints reported with this failure mode in the past three years.The devices were intended for use in treatment.As no samples were returned a product evaluation could not be carried out.Dressings should be periodically monitored to ensure proper catheter placement.If the dressing becomes wet or the integrity is compromised, catheter placement may be impacted.We have not been able to confirm a relationship between the event and the device or identify a definitive root cause.No further actions by smith and nephew are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN IV3000
Type of Device
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key10231063
MDR Text Key198080480
Report Number8043484-2020-01530
Device Sequence Number1
Product Code KGX
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKN05003901
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/10/2020
Initial Date FDA Received07/05/2020
Supplement Dates Manufacturer Received10/07/2020
Supplement Dates FDA Received10/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-