• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER Back to Search Results
Model Number CV-180
Device Problems Circuit Failure (1089); No Display/Image (1183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Since the subject device has not been returned to omsc for the evaluation, the exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed from the user facility that the image of the subject device couldn't be displayed on the monitor before the unspecified procedure.The user replaced the system including the subject device to another system and completed the procedure.The service department of olympus (b)(4) (oekg) checked the subject device and identified it was occurred due to the print circuit board failure.There was no report of patient injury associated with this event.The user did not provide the other detailed information.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Since the subject device was not returned to olympus medical systems corp.(omsc), it could not be investigated.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.However based on the report of the service department of olympus europe se & co.Kg (oekg), omsc surmised there was the possibility this phenomenon was attributed to dx printed circuit board failure which is equipped with the compout connector due to the long-term use of the subject device since manufacturing in february 2007.If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10231405
MDR Text Key223653826
Report Number8010047-2020-03905
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberCV-180
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/11/2020
Initial Date FDA Received07/06/2020
Supplement Dates Manufacturer Received08/31/2020
Supplement Dates FDA Received09/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-