Details of complaint: the biomedical engineer (bme) reported that the telemetry transmitter started to heat up and got too hot to touch or hold.No patient harm was reported.Service requested / performed: evaluation / repair: the reported issue was not duplicated by nihon kohden repair center (nk rc).During the evaluation, corrosion and residue was found on the battery contacts and case.There was also fluid intrusion to the main board.Investigation summary: as per the hazard analysis document 0704-905438f, the overheating of the battery compartment could be caused due to a number of reasons, such as: ·the battery is inserted in a wrong polarity causing overcurrent ·battery is inserted forcibly, causing fire due to short-circuit of +/- terminals ·short-circuit between the +/- battery terminals ·user does not handle nimh secondary battery and a battery charger properly, causing a short-circuit of the +/- terminals investigation determined that the reported issue could have been caused due to incorrect usage and maintenance of nk device.No further investigation is needed through the capa process.
|