SMITH & NEPHEW, INC. CONTROL UNIT DYONICS 25; ARTHROSCOPE Back to Search Results
Model Number 7211010
Device Problem Insufficient Information (3190)
Patient Problem No Information (3190)
Event Date 06/05/2020
Event Type  malfunction  
Event Description
It was reported that, during surgery set up and prior being used, the "control unit dyonics 25" was not taking the cassette.The procedure was successfully completed without delay using a back-up device.No patient injuries or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The reported device, intended for use in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.Complaint of cassette insertion failure could not be confirmed.Three different types of cassettes were inserted into the pump with no problem.All 3 cassettes locked into place.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.The device has been in service for 7 years, thus any malfunction presented at this point in time would not be related to the manufacture of the product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTROL UNIT DYONICS 25
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key10242684
MDR Text Key197976936
Report Number1643264-2020-00491
Device Sequence Number1
Product Code HRX
UDI-Device Identifier03596010543554
UDI-Public03596010543554
Combination Product (y/n)N
PMA/PMN Number
K051326
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 08/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7211010
Device Catalogue Number7211010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2020
Initial Date Manufacturer Received 06/16/2020
Initial Date FDA Received07/07/2020
Supplement Dates Manufacturer Received08/14/2020
Supplement Dates FDA Received08/18/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-