Model Number 2217-50-041
Device Problems Device-Device Incompatibility (2919); Material Twisted/Bent (2981)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/25/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the liner impactor has cross threaded on cup impactor.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint; impactor is cross threaded on cup impactor.The impactor tip was able to be disassembled from the impactor.The threads of the impactor tip exhibited deformation (damage) that is not possible with the intended use of the handle.When the damaged impactor was attached to the impactor head, damage to the impactor head internal threads occurred and made disassembly difficult.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.H10 additional narrative: corrected: h3.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d10.Corrected: h3.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN STRAIGHT CUP IMPACTOR
Type of Device
HIP INSTRUMENTS : IMPACTORS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10254665
MDR Text Key198305098
Report Number1818910-2020-15483
Device Sequence Number1
Product Code MRA
UDI-Device Identifier10603295098980
UDI-Public10603295098980
Combination Product (y/n)N
PMA/PMN Number
P070026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 06/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2217-50-041
Device Catalogue Number221750041
Device Lot NumberNB122043
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2020
Initial Date Manufacturer Received 06/25/2020
Initial Date FDA Received07/09/2020
Supplement Dates Manufacturer Received06/25/2020
08/07/2020
Supplement Dates FDA Received07/13/2020
08/14/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-