• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. FA COBLATOR II CONTROLLER (120V); ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. FA COBLATOR II CONTROLLER (120V); ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number 13546-01
Device Problem Defective Component (2292)
Patient Problem No Patient Involvement (2645)
Event Date 06/18/2020
Event Type  malfunction  
Event Description
It was reported that the foot pedal plug on the generator was loose and not connecting securely.No case involved.Therefore, there was no patient involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H10: h3, h6: the reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A review of the device history records show there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found related failures; this failure mode will be trended to assess for any necessary corrective actions.A visual inspection observed that the footpedal port is loose and spins in place, the lid and bezel are damaged, the overlay is dented, and the unit is missing a foot.A functional evaluation revealed no issues.The unit was opened and found no issues.The complaint was verified.Factors, unrelated to the design or manufacture of the device which could have contributed to the complaint event, include misalignment of the connector when connecting to the unit receptacle in which excessive force or twisting could cause damage.No containment or corrective actions are recommended at this time.There are no indications to suggest that the device/product did not meet specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FA COBLATOR II CONTROLLER (120V)
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key10254667
MDR Text Key198343361
Report Number3006524618-2020-00458
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
PMA/PMN Number
K030108
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 08/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number13546-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/17/2020
Initial Date Manufacturer Received 06/12/2020
Initial Date FDA Received07/09/2020
Supplement Dates Manufacturer Received08/07/2020
Supplement Dates FDA Received08/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-