RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problems
Mechanical Problem (1384); Physical Resistance/Sticking (4012)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 09jul2020.
|
|
Event Description
|
The customer reported a ventilator that will not go into standby mode.This event was resolved via conversation between the customer and the manufacturer.There was no on-site evaluation by the manufacturer.This event was reported to have occurred during clinical use.The customer reported that they were observing too much resistance to flow.It was surmised by the customer that the filters being used may be contributing to an out-of-specification resistance level, and may not be a recommended filter for the ventilator.Changes to the set-up of the circuit on the ventilator were discussed, but it was concluded that further repair and replacement were not required.
|
|