H10: after further assessment of the information gathered by the manufacturer, it was identified that this event should be re-evaluated for mdr reporting.The original information stated that, during a tka procedure, it was found that both femur and tibial cutting blocks included in vis cut guide jbcs kit rt did not fit at all.Surgeon completed case with standard instrumentation.A delay no greater than 30 min was reported.No patient harm was reported to occur as a consequence of this malfunction.However, after further clarification and information received, it was determined that the present complaint does not meet the criteria to be reportable nor to be a complaint, because the two devices involved: the femoral and the tibial cutting blocks, are included in the vis cut guide jbcs kit rt.Therefore, the report already submitted (report number: 9613369-2020-00138) covers the malfunction of both devices.
|