This problem was caused by lot to lot variation of a raw material, serum.In the acceptance inspection, an accelerated stability test is done to evaluate the stability of a new lot of serum.The problematic serum passed the test but deteriorated faster than expected.Risk assessment: for diagnosis of pregnancy, ifu shows non-pregnant reference value is < 0.5 miu/ml and it was reported that the hcg concentration is 5 to 50 miu/ml within the 1st week of pregnancy.It is also reported that hcg concentration doubles every 48 hr.During early pregnancy.Usually pregnancy is diagnosed in conjunction with ultrasonography.After hcg reached 1000 to 2000 miu/ml, the gestational sac could be observed by ultrasonography.Because of above mentioned dynamic change of hcg concentration, 20% higher hcg value do not seems to result in health hazard.
|
A customer in (b)(6) reported that the measurement values of control (tosoh multi-control) were around or exceed upper value of assigned range.Internal investigation found poor stability of the calibrator set.Because of this, if this calibrator set were used around its expiration date, test results of patient samples as well as controls could be 20% higher than original value.
|