• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. FLEXIBLE SHAFT W/CIR CONNECTOR; PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. FLEXIBLE SHAFT W/CIR CONNECTOR; PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED Back to Search Results
Catalog Number 71118200
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 06/18/2020
Event Type  malfunction  
Event Description
It was reported that the end of the reamer was broken.No surgery or patient involved.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.A visual inspection confirms that the flexible reamer shaft fractured within the shaft via mechanical overload.If sufficient offset bending or torsional forces are applied to the reamer shaft during use that exceeds the strength of the material, a mechanical overload fracture can occur.The device was manufactured in 2013.This device shows significant signs of wear/usage.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the device history records for the listed batch did not reveal any deviation from the standard manufacturing processes.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIBLE SHAFT W/CIR CONNECTOR
Type of Device
PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10262340
MDR Text Key198481964
Report Number1020279-2020-03137
Device Sequence Number1
Product Code HSD
Combination Product (y/n)N
PMA/PMN Number
K121714
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 09/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71118200
Device Lot Number13EM00323
Initial Date Manufacturer Received 06/18/2020
Initial Date FDA Received07/13/2020
Supplement Dates Manufacturer Received08/24/2020
Supplement Dates FDA Received09/02/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-