• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. REPAIR, STEM IMPACTOR ROD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. REPAIR, STEM IMPACTOR ROD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 71364011R
Device Problem Material Integrity Problem (2978)
Patient Problem No Patient Involvement (2645)
Event Date 06/18/2020
Event Type  malfunction  
Event Description
It was reported that the stem impactor rod was found stripped.No procedure-related.
 
Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection of the returned stem impactor rod confirms the tip threads are damaged.The device is manufactured in 2003.The device exhibits signs of extreme wear and use.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REPAIR, STEM IMPACTOR ROD
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10268731
MDR Text Key198676994
Report Number1020279-2020-03172
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 08/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71364011R
Initial Date Manufacturer Received 06/18/2020
Initial Date FDA Received07/13/2020
Supplement Dates Manufacturer Received06/18/2020
Supplement Dates FDA Received08/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-