• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS POWERPORT SLIM IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, KIT, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVSCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS POWERPORT SLIM IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, KIT, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVSCULAR Back to Search Results
Model Number 1718000
Device Problems Obstruction of Flow (2423); Material Deformation (2976); Appropriate Term/Code Not Available (3191)
Patient Problems Inflammation (1932); Pain (1994)
Event Type  Injury  
Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Expiry date (08/2020).
 
Event Description
It was reported that post port placement, the patient allegedly experienced pain along the catheter track and inflammation in the neck.It was further reported that the catheter has compression at the clavicle.Reportedly, the port only works when the patient lays down.
 
Event Description
It was reported that post port placement, the patient allegedly experienced pain along the catheter track and inflammation in the neck.It was further reported that the catheter has compression at the clavicle.Reportedly, the port only works when the patient lays down.
 
Manufacturer Narrative
H10: manufacturing review: a lot history review was performed.This is the only complaint to date for this lot number.Therefore, a device history record (dhr) review is not required.Investigation summary: the investigation is inconclusive for the reported catheter compression and swelling issue, as the device was not returned for evaluation.Based upon the available information, the definitive root cause could not be determined.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT SLIM IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, KIT, 8F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVSCULAR
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key10269344
MDR Text Key198689817
Report Number3006260740-2020-02412
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741026782
UDI-Public(01)00801741026782
Combination Product (y/n)N
PMA/PMN Number
K072215
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1718000
Device Catalogue Number1718000
Device Lot NumberREDQ2925
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/19/2020
Initial Date FDA Received07/14/2020
Supplement Dates Manufacturer Received11/11/2020
Supplement Dates FDA Received11/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-