The device, used in treatment, was returned for evaluation.A visual inspection of the returned evos fixed handle confirms the tip of the device is broken off.The broken piece was not returned with the device.This instrument was manufactured in 2017.This device shows significant signs of wear/ usage.A medical investigation was conducted and per the complaint details, during the ankle orif, the screwdriver tip snapped; however, all pieces were retrieved and the procedure was successfully completed with a backup device within a 0-30 minute surgical procedure.Reportedly, no additional interventions or x-rays were required.The patient impact beyond the reported event and the modified surgical procedure using a backup device could not be determined, although no patient injury was reported.No further medical assessment is warranted at this time.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of risk management files found that the reported failure was documented appropriately.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|