SMITH & NEPHEW, INC. ARTICULAR SURFACE PROVISIONAL SZ 4 9MM; TEMPLATE Back to Search Results
Catalog Number 00584102409
Device Problem Material Fragmentation (1261)
Patient Problem No Patient Involvement (2645)
Event Date 06/24/2020
Event Type  malfunction  
Event Description
It was reported that the articular surface provisional sz 4 9mm was worn, with broken sides and would not fully insert.This was found during set up / inspection.No patient involvement.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned articular surface provisional confirms multiple scratches, burrs and deep gouges in the plastic.There is also a crack on the top of the device.The device shows significant signs of wear/ usage.This instrument was manufactured in 2016.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.The device is a reusable instrument that can be exposed to numerous surgeries and cleaning cycles.As plastics are vulnerable and crack may have initiated during use and possible causes could be due to the heating and cooling associated with autoclaving or prolonged use.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE PROVISIONAL SZ 4 9MM
Type of Device
TEMPLATE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10273903
MDR Text Key198878124
Report Number1020279-2020-03203
Device Sequence Number1
Product Code HWT
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 09/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number00584102409
Device Lot Number63416090
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 06/24/2020
Initial Date FDA Received07/14/2020
Supplement Dates Manufacturer Received08/11/2020
Supplement Dates FDA Received09/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-