• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ARTICULAR SURFACE PROVISIONAL SZ 4 8MM; TEMPLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ARTICULAR SURFACE PROVISIONAL SZ 4 8MM; TEMPLATE Back to Search Results
Model Number 00584102408
Device Problem Material Fragmentation (1261)
Patient Problem No Patient Involvement (2645)
Event Date 06/24/2020
Event Type  malfunction  
Event Description
It was reported that the poly pieces were found worn, with broken sides and will not insert fully.This was found during set up / inspection.No surgery or patient involvement.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.A visual inspection of the returned device confirms a multiple scratches, burrs and deep gouges in the plastic.The device shows extreme signs of wear/usage.The lot number can¿t be read on the device.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.The device is a reusable instrument that can be exposed to numerous surgeries and cleaning cycles.As plastics are vulnerable and crack may have initiated during use and possible causes could be due to the heating and cooling associated with autoclaving or prolonged use.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE PROVISIONAL SZ 4 8MM
Type of Device
TEMPLATE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10279833
MDR Text Key199044143
Report Number1020279-2020-03232
Device Sequence Number1
Product Code HWT
UDI-Device Identifier00885556619506
UDI-Public00885556619506
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 08/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number00584102408
Device Catalogue Number00584102408
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/21/2020
Initial Date Manufacturer Received 06/24/2020
Initial Date FDA Received07/15/2020
Supplement Dates Manufacturer Received08/06/2020
Supplement Dates FDA Received08/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-