ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752201
Device Problem Obstruction of Flow (2423)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/15/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that the infusion sleeve folded up on itself resulting in obstruction during a procedure.The product was replaced and procedure completed with no patient harm.
 
Manufacturer Narrative
There were two fluid management system finished good lots associated with the custom pak.It was not known which lot was used for this event.A review of the device history record traceable to the reported lot numbers indicates that the product was processed and released according to the product¿s acceptance criteria for both lots.All lots are verified that all required inspections have been performed and all acceptance criteria are met prior to release.No sample has been returned for evaluation; therefore, the condition of the product could not be verified.The root cause of the customer's complaint could not be established as a sample has not been received and the condition of the product could not be verified.Without analysis of the sample, it is not possible to isolate the root cause.As the root cause is unknown, the relationship, if any, of the device to the reported incident cannot be determined.The root cause for this complaint is not known, therefore, specific action with regards to this complaint cannot be taken.All lots are verified that all required inspections have been performed and all acceptance criteria are met prior to release.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.Consumables manufacturing management has been made aware of this complaint through the consumer affairs review meeting.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
Additional information indicates that the surgery was performed on left eye.The manufacturer internal reference number is: (b)(4).
 
Event Description
Additional information indicates that the surgery was performed on left eye.
 
Manufacturer Narrative
Correction provided in d.4.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
MDR Report Key10285804
MDR Text Key199873514
Report Number1644019-2020-00374
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup,Followup,Followup
Report Date 08/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065752201
Device Lot NumberASKU
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/16/2020
Initial Date FDA Received07/16/2020
Supplement Dates Manufacturer Received07/27/2020
07/27/2020
08/21/2020
Supplement Dates FDA Received08/10/2020
08/20/2020
08/31/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CENTURION VISION SYSTEM; CUSTOM PAK; OZIL HANDPIECE
-
-