G4: 20oct2020.B4: 26oct2020.V60 power supply was received for failure evaluation.Visual inspection revealed no evidence of damage or contamination.A failure investigation (fi) technician installed the v60 power supply into a fi ventilator to duplicate the reported problem.During the unit testing, the fi technician identified that q1 (500v single n-channel hexfet power mosfet) leads 1-3 shorted and cr5 (turbo 2 ultrafast high voltage rectifier) leads 1-2 shorted.The customer complaint was verified.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|