• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ETS FLEX ARTICNG LNR CUTR 45MM; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ETS FLEX ARTICNG LNR CUTR 45MM; STAPLE, IMPLANTABLE Back to Search Results
Model Number ATS45
Device Problem Crack (1135)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/03/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # u5ay4k.Investigation summary: the analysis results found that the ats45 device was received with the firing mechanism damaged and two tr45w cartridges loaded in the device.The reload (b) was received unfired.The reload (c) was received partially fired 1/10 with the reload lock out spring damaged.No functional test could be performed due to the condition of the device.The device was disassembled to verify the condition of the internal components and the firing trigger teeth were found broken.A manufacturing record evaluation was performed for the finished device and the manufacturing criteria were met prior to the release of this lot/ batch.The damage to the reload lockout spring is consistent with damage observed when the firing cycle is started, interrupted, released, and restarted.When firing the device, make sure that the firing stroke is completed.Do not partially fire the device.Fire the device by squeezing the firing trigger completely until it rests on the closing trigger.Once the firing cycle has been initiated, it must be completed.If re-initiation of firing is resumed, the device will lockout.Firing through the lockout mechanism will break the device.Please reference the instruction for use for more information.It should be noted that a 100% inspection takes place during manufacturing to ensure the device meets the require specifications.In addition, a sample of the batch is inspected at (b)(4).
 
Event Description
It was reported that during a laparoscopic appendectomy, on the second firing, a 'crack' was heard and the device locked out.There was no difficulty opening the device and the first firing went fine.A similar device was used to complete the case with no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETS FLEX ARTICNG LNR CUTR 45MM
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 00969
*   00969
6107428552
MDR Report Key10289167
MDR Text Key199995317
Report Number3005075853-2020-03627
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10705036001072
UDI-Public10705036001072
Combination Product (y/n)N
PMA/PMN Number
K020779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberATS45
Device Catalogue NumberATS45
Device Lot NumberU93V4M
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2020
Initial Date Manufacturer Received 06/19/2020
Initial Date FDA Received07/17/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/13/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-