RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Computer Operating System Problem (2898)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 18jul2020.
|
|
Event Description
|
The customer reported that during preventive maintenance (pm) service, the unit will not run on battery.It shuts down when alternating current power is unplugged.The customer contacted product support and requested for service repair.The biomed, swapped in a known good battery with the same results.The customer reported that the unit was not in use on a patient.The field service engineer (fse) was able to verify the issue.The unit will not run if unplugged.The power management board for potential power management board failure.The fse replaced the power management board and performed required tests per the service manual.All tests passed.
|
|