Model Number V60
Device Problem Computer Operating System Problem (2898)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 20jul2020.
 
Event Description
The customer contacted technical support (ts) stating that the unit would not go into standby mode.The customer reported there was no patient involvement at the time the issue was discovered.
 
Manufacturer Narrative
G4: 03jan2021.B4: 19jan2021.The manufacturer's product service engineer (pse) was dispatched to the site to evaluate the unit and confirmed that the flow sensor was defective.The pse replaced the flow sensor assembly, air & o2.The ventilator passed all testing and operated within the manufacturing specifications.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
MDR Report Key10299350
MDR Text Key200397737
Report Number2031642-2020-02423
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 06/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/26/2020
Initial Date FDA Received07/20/2020
Supplement Dates Manufacturer Received06/26/2020
Supplement Dates FDA Received01/19/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/15/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-