SMITH & NEPHEW, INC. LIGHT SOURCE 500XL XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
|
Back to Search Results |
|
Model Number 72200568 |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/01/1901 |
Event Type
malfunction
|
Event Description
|
It was reported that the fan of the light source was not working and device overheated.It is unknown whether the event happened during surgery and if there was a patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The reported device, intended for use in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.A relationship, if any, between the subject device and the reported event could not be determined.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
|
|
Manufacturer Narrative
|
The reported device, intended for use in treatment, was received for evaluation.A visual inspection was performed on the exterior of product and no physical damage was observed.There was a relationship found between the returned device and the reported incident.The complaint of fan malfunction and overheating was confirmed.The cooling fan for the lamp shuts down when unit is powered up causing overheating and errors.Fan malfunction is caused by a defective electronic component.The complaint investigation has concluded the cause of the failure to be a defective electronic component.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
|
|
Search Alerts/Recalls
|
|
|