BAXTER HEALTHCARE CORPORATION CLEARLINK BLOOD RECIPIENT SET; SET, BLOOD TRANSFUSION Back to Search Results
Catalog Number 4C8723
Device Problem Component Missing (2306)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
This event occurred during the unspecified date of (b)(6) 2020.The device was received for evaluation.Visual inspection was performed using the naked eye which observed that the roller clamp was missing.The reported condition was verified.The cause of the condition is supplier manufacturing related.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
During evaluation a clearlink system y-type blood/solution set was returned with the roller clamp missing.This was identified prior to use.There was no patient involvement.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK BLOOD RECIPIENT SET
Type of Device
SET, BLOOD TRANSFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key10302620
MDR Text Key199715052
Report Number1416980-2020-04143
Device Sequence Number1
Product Code BRZ
UDI-Device Identifier00085412049076
UDI-Public(01)00085412049076
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K993120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 07/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number4C8723
Device Lot NumberR20B18095
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/17/2020
Initial Date Manufacturer Received 07/07/2020
Initial Date FDA Received07/21/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/19/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-