Event description: customer found "bottom broke off during a case." the device was returned, the complaint was confirmed, the device was fractured near the braze joint at the proximal end.Failure in this manner is often a result of the application of excessive torque during use which causes wear against the endoscope at the proximal end of the instrument channel.The device ifu (spl-ifu rev am) warns on page 22, "the probe is fragile.It is critical that the surgeon does not bend or torque the probes against the endoscope during the procedure.There is no need to rotate the probe or transducer; it will not improve fragmentation or stone clearance." olympus will continue to monitor the field performance of this device.A review of the device history record was completed and it was confirmed that there were no abnormalities, special adoptions, or variations in the manufacturing process.
|