The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction centrifuge bowl leak/break.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot h168 was conducted.There were no non-conformances associated with this lot.This lot met all release requirements.A review of kit lot h168 shows no trends.Trends were reviewed for complaint categories, alarm #7: blood leak? (centrifuge chamber), centrifuge bowl leak/break.No trends were detected for each complaint category.The customer returned photographs for evaluation.The supplied photographs verify the centrifuge bowl leaked as blood splatter is seen on the centrifuge chamber walls.The bowl appears to be intact and still locked into the bowl holder of the cellex instrument.The photographs show a crack running across the side of the centrifuge bowl.A material trace of the bowl assembly and its components used to build lot h168 found no related non-conformances.A device history record review of kit lot h168 did not identify any related non-conformances and this kit lot had passed all lot release testing.The cause of the centrifuge bowl leak was most likely due to the crack in the outer centrifuge bowl.The root cause of the crack in the centrifuge bowl could not be determined based on the available information.No further action is required at this time.This investigation is now complete.(b)(4).
|