ZIMMER BIOMET, INC. BIOMET CC I-BEAM TRAY 71MM; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problems Loss of or Failure to Bond (1068); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problems Unspecified Infection (1930); Pain (1994); Tissue Damage (2104); Blood Loss (2597)
Event Date 09/10/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated report: 0001825034-2020-02902.0001825034-2020-02904.Medical product: vngd ps tib brg 10x71/75mm item# 183640 lot# 040510.Van ps open intl fem-lt 70 item# 183132 lot# j3880066.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that a patient underwent a phase one revision approximately one year and seven weeks post implantation due to infection.An arthroscopy was performed four months earlier.The tibial component was found to be loose as a consequence of the infection.All components were removed and antibiotic cement spacers were placed.A picc line was placed and the patient received iv antibiotics.Three months and eleven days later the patient was re-implanted with a new left total knee arthroplasty.There is no additional information at this time.
 
Manufacturer Narrative
This follow-up is being submitted to relay additional information.Updated: b4, b5, d4, g4, g7, h1, h2, h3, h4, h6, and h10 the reported event of deep infection greater than ninety days occurred post implantation.During the investigation process a review of the sterile certifications were reviewed and found to be conforming with no applicable deviations.Devices were verified to have gone through acceptable sterilization process following iso/aami/astm & eu published guidelines.There are multiple factors that may contribute to an infection that are outside the control of zimmer biomet, such as external factors, i.E.Hospital/surgical environment, provider related risk factors, and/or patient comorbidities/risk factors.As there are no indications of a product or process issues identified affecting implant safety or effectiveness, therefore implanted products are not identified as the source or contributing to the reported infection.Noted events of tibial component loosening, blood loss and postop anemia are all secondary to the infection, as noted by the surgeon in the medical records.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOMET CC I-BEAM TRAY 71MM
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10326767
MDR Text Key200598113
Report Number0001825034-2020-02903
Device Sequence Number1
Product Code JWH
Combination Product (y/n)N
PMA/PMN Number
K142933
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 12/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number141223
Device Lot NumberJ6061624
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/02/2020
Initial Date FDA Received07/27/2020
Supplement Dates Manufacturer Received12/03/2020
Supplement Dates FDA Received12/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Weight81
-
-