• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. ACHIEVE ADVANCE MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. ACHIEVE ADVANCE MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number 2ACH20
Device Problems Detachment of Device or Device Component (2907); Material Twisted/Bent (2981)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/26/2020
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the mapping catheter 2ach20 with lot 11138639, was returned and analyzed.Visual inspection showed that the pebax tube was detached and the loop was damaged as it was stuck inside the luer.Additional imaging showed that four electrodes were stuck inside the luer of the catheter.In conclusion, the mapping catheter failed inspection due to loop damage and detachment.If information is provided in the future, a supplemental report will be issued.
 
Event Description
After a completed case, the balloon catheter and mapping catheter subsequently tested out of specification per the manufacturer's investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACHIEVE ADVANCE MAPPING CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key10338540
MDR Text Key200812507
Report Number2182208-2020-01382
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162892
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/07/2021
Device Model Number2ACH20
Device Catalogue Number2ACH20
Device Lot Number11138639
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2020
Initial Date Manufacturer Received 07/18/2020
Initial Date FDA Received07/29/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/08/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
BALLOON CATHETER 2AF283
-
-