• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JOURNEY TIBIA BASE NP RT SZ 3; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JOURNEY TIBIA BASE NP RT SZ 3; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 74022213
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Injury (2348); Joint Swelling (2356)
Event Date 09/12/2018
Event Type  Injury  
Event Description
It was reported that cellulitis on the right knee and leg had presented.Patient developed redness and swelling of the leg.Admitted to hospital for antibiotics.Eventually resolved without long term problems.Home physical therapy and medication were prescribed for treatment.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical evaluation concluded that the journey ii cr clinical study reports a patient developed cellulitis with redness and swelling of the right knee and leg.The patient was admitted to hospital for antibiotics.Eventually the event resolved without long term problems.Home physical therapy and medication were also prescribed.The clinical study report forms were provided for review, however they did not reveal a root cause for the cellulitis event.It has been communicated that no further medical documentation would be forthcoming.Smith and nephew has not received adequate materials to fully evaluate the complaint, but if additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of sterilization documents indicated that the product was sterilized according to sterilization release documentation.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use documents revealed this failure mode and potential harm was previously identified.Infection, a potential complication associated with any surgery, can occur and possible causes could include but not limited to contamination, patient reaction, and post-operative healing issue.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved or product information, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOURNEY TIBIA BASE NP RT SZ 3
Type of Device
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10342328
MDR Text Key200964112
Report Number1020279-2020-03573
Device Sequence Number1
Product Code JWH
UDI-Device Identifier03596010554277
UDI-Public03596010554277
Combination Product (y/n)N
PMA/PMN Number
K042515
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 07/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number74022213
Device Catalogue Number74022213
Device Lot Number18DM22472
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/06/2020
Initial Date FDA Received07/29/2020
Supplement Dates Manufacturer Received07/13/2021
Supplement Dates FDA Received07/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age69 YR
Patient Weight82
-
-