The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction drive tube leak/break.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot h374 was conducted.There were no non-conformances associated with this lot.This lot met all release requirements.A review of kit lot h374 shows no trends.Trends were reviewed for complaint categories, alarm #7: blood leak (centrifuge chamber), and drive tube leak/break.No trends were detected for each complaint category.The complaint kit and photographs were returned for evaluation.The provided photographs verify a blood leak occurred as blood spray is seen in a band around the centrifuge chamber walls.The drive tube and centrifuge bowl are intact and still secured into the drive tube retainers and centrifuge bowl holder.Examination of the received kit found dried blood on the drive tube component of the kit.The drive tube and centrifuge bowl were pressure tested to check for leaks and no leaks were identified.The returned kit was installed onto a test instrument for dynamic testing.A dyed solution was pumped into the centrifuge bowl and drive tube and the centrifuge was spun at 2400 rpm for 10 minutes.The leak sensor did not alarm during testing and inspection of the centrifuge chamber did not identify any leaks.Testing of the returned kit did not identify a potential leak site.A material trace of the bowl assembly and its components used to build lot h374 found no related non-conformances.A device history record (dhr) review did not result in any related non-conformances and this kit lot passed all lot release testing.The cause of the alarm #7: blood leak? (centrifuge chamber) was most likely the blood leak in the centrifuge chamber.The root cause of the drive tube leak could not be determined based on the available information.No further action is required at this time.This investigation is now complete.(b)(4).
|