The patient was undergoing a thrombectomy procedure using a penumbra system aspiration pump max 110 (pump max).During the procedure, the hospital staff noticed that the pump max did not generate vacuum pressure.An attempt was made to troubleshoot the pump max but was unsuccessful.It was reported that the regulatory knob was loose, spinning and not catching on the dial.Subsequently, not allowing the vacuum pressure to be adjusted upward.Therefore, the pump max was disconnected.The physician decided to abandon the thrombus aspiration and opted to angioplasty.There was no report of an adverse effect to the patient.
|
Results: dried blood was observed on the pump housing.The regulator knob was undamaged.Conclusions: evaluation of the returned pump max revealed that blood was aspirated inside its vacuum assembly.If the aspiration tubing is connected directly to the vacuum inlet instead of the canister, supplied by penumbra, blood will likely be aspirated into the vacuum assembly.If fluid is aspirated into the vacuum assembly, the pump may not function properly.The regulator knob was visually inspected, and no damage was observed.Therefore, the reported regulator knob damage could not be confirmed.Penumbra pumps are visually inspected and functionally tested during incoming quality inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|