• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX I.C. FOLEY CATHETER; FOLEY CATHETER (LATEX)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX I.C. FOLEY CATHETER; FOLEY CATHETER (LATEX) Back to Search Results
Model Number 0165SI16
Device Problems Partial Blockage (1065); Obstruction of Flow (2423)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event Description
It was reported that the foley catheters kept getting clogged frequently, which resulted in an increase in changing the foley.
 
Event Description
It was reported that the foley catheters kept getting clogged frequently, which resulted in an increase in changing the foley.
 
Manufacturer Narrative
The reported event was inconclusive as no sample was returned.The potential root cause for this failure mode could be user related (example: salt accumulation)/ block drainage lumen /no drainage eye.The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "caution: this product contains natural rubber latex which may cause allergic reactions.To deflate catheter balloon: gently insert a luer slip syringe in the catheter valve.Never use more force than is required to make the syringe "stick" in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all recommended inflation capacities 3cc balloon: use 5cc sterile water 5cc balloon: use 10cc sterile water 30cc balloon: use 35cc sterile water warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with applicable local, state, and federal laws and regulations.Catheters should be replaced in accordance with the cdc guideline, "guideline for prevention of catheter-associated urinary tract infection." at the onset or first signs of a urinary tract infection, catheter encrustation, or any other catheterrelated adverse effect, the catheter should be replaced.Do not exceed recommended capacities.Sterile: unless package is opened or damaged.Do not use if package is opened or damaged.Single use only.Do not reuse.Do not resterilize.For urological use only.Valve type: use luer slip syringe.Do not use needle.Caution: do not aspirate urine through drainage funnel wall.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Visually inspect the product for any imperfections or surface deterioration prior to use.Warning: on catheter, do not use ointments or lubricants having petrolatum base.They will damage latex and may cause the balloon to burst.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX I.C. FOLEY CATHETER
Type of Device
FOLEY CATHETER (LATEX)
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key10350071
MDR Text Key202515728
Report Number1018233-2020-04884
Device Sequence Number1
Product Code MJC
UDI-Device Identifier00801741016752
UDI-Public(01)00801741016752
Combination Product (y/n)N
PMA/PMN Number
K040658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,distributor,other
Type of Report Initial,Followup
Report Date 09/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number0165SI16
Device Catalogue Number0165SI16
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/09/2020
Initial Date FDA Received07/31/2020
Supplement Dates Manufacturer Received08/24/2020
Supplement Dates FDA Received09/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-