Model Number 0570-0313 |
Device Problems
Erratic or Intermittent Display (1182); No Display/Image (1183)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/30/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device return is anticipated; however, at the time of the report the device has not been received by verathon.Verathon continues to investigate the reported event and a supplemental report will be submitted in accordance with 21 cfr 803.56 when additional information becomes available.Upon review of the device history for serial number (b)(4), it was determined that the device was manufactured on 19 jun 2017 and is past the two (2) year expected product life as outlined in the glidescope operations and maintenance manual (omm).
|
|
Event Description
|
The customer reported that during a patient procedure, using a glidescope avl video baton 3-4, the image was going in and out when the cable was moved.No delay in the procedure, use of a backup device, or harm to the patient or user was reported.
|
|
Manufacturer Narrative
|
The glidescope avl video baton 3-4 was returned to verathon for evaluation.A verathon technical service representative evaluated the returned glidescope avl video baton 3-4 and confirmed the poor image quality issue.The technical service representative attributed the image issue to cable failure.The glidescope avl video baton 3-4 was scrapped and a replacement was sent to the customer.Corrective action is not required at this time.Verathon will continue to monitor for trends.
|
|