• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VERATHON MEDICAL ULC GLIDESCOPE AVL VIDEO BATON 3-4; LARYNGOSCOPE, RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VERATHON MEDICAL ULC GLIDESCOPE AVL VIDEO BATON 3-4; LARYNGOSCOPE, RIGID Back to Search Results
Model Number 0570-0313
Device Problems Erratic or Intermittent Display (1182); No Display/Image (1183)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 06/30/2020
Event Type  malfunction  
Manufacturer Narrative
The device return is anticipated; however, at the time of the report the device has not been received by verathon.Verathon continues to investigate the reported event and a supplemental report will be submitted in accordance with 21 cfr 803.56 when additional information becomes available.Upon review of the device history for serial number (b)(4), it was determined that the device was manufactured on 19 jun 2017 and is past the two (2) year expected product life as outlined in the glidescope operations and maintenance manual (omm).
 
Event Description
The customer reported that during a patient procedure, using a glidescope avl video baton 3-4, the image was going in and out when the cable was moved.No delay in the procedure, use of a backup device, or harm to the patient or user was reported.
 
Manufacturer Narrative
The glidescope avl video baton 3-4 was returned to verathon for evaluation.A verathon technical service representative evaluated the returned glidescope avl video baton 3-4 and confirmed the poor image quality issue.The technical service representative attributed the image issue to cable failure.The glidescope avl video baton 3-4 was scrapped and a replacement was sent to the customer.Corrective action is not required at this time.Verathon will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE AVL VIDEO BATON 3-4
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
MDR Report Key10350549
MDR Text Key202533460
Report Number9615393-2020-00163
Device Sequence Number1
Product Code CCW
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 06/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0570-0313
Device Catalogue Number0570-0313
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2020
Initial Date Manufacturer Received 06/30/2020
Initial Date FDA Received07/31/2020
Supplement Dates Manufacturer Received08/26/2020
Supplement Dates FDA Received09/02/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-