• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. REF CUP POSITIONER/IMPACTOR; PRSTHESS, HP, SMI-CNSTRIND, MTL/PLYMR, CMNTD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. REF CUP POSITIONER/IMPACTOR; PRSTHESS, HP, SMI-CNSTRIND, MTL/PLYMR, CMNTD Back to Search Results
Catalog Number 71362299
Device Problem Material Deformation (2976)
Patient Problem No Patient Involvement (2645)
Event Date 07/07/2020
Event Type  malfunction  
Event Description
It was reported that the threads of the impactor are damaged.No case involved.
 
Manufacturer Narrative
The device, intended use in treatment, was returned for evaluation.A visual inspection confirmed the threads are damaged on the ref cup positioner/impactor.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REF CUP POSITIONER/IMPACTOR
Type of Device
PRSTHESS, HP, SMI-CNSTRIND, MTL/PLYMR, CMNTD
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10350876
MDR Text Key201211746
Report Number1020279-2020-03598
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71362299
Device Lot Number15AM09821
Initial Date Manufacturer Received 07/07/2020
Initial Date FDA Received07/31/2020
Supplement Dates Manufacturer Received08/18/2020
Supplement Dates FDA Received09/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-