• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO FIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO FIBERSCOPE Back to Search Results
Model Number URF-P7
Device Problems Fluid/Blood Leak (1250); Image Display Error/Artifact (1304)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
This supplemental report is submitted to provide the results of the legal manufacturer¿s investigation and device history record (dhr) review.The dhr for the subject device was reviewed and it was verified the device was manufactured in accordance with documented specifications.The legal manufacturer performed an investigation and confirmed that the cause of breakage of the bending tube was likely due to user handling or technique.
 
Manufacturer Narrative
The scope was returned to olympus and evaluated.The reported problem was confirmed.A leak was found at the a-rubber.In addition, excessive broken fibers and a broken skeleton ring were found.Based on the results of the device evaluation and similar reported complaints, the most likely cause of the reported event can be attributed to user handling or technique.The device instructions for use contains the following statements: caution: "do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist." "do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damaged.".
 
Event Description
A user facility reported to olympus "leaking and black spots on image." there was no patient injury or harm, associated with the problem, reported to olympus.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Type of Device
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10351218
MDR Text Key224556976
Report Number8010047-2020-05093
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-P7
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2020
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/07/2020
Initial Date FDA Received07/31/2020
Supplement Dates Manufacturer Received09/30/2020
Supplement Dates FDA Received10/26/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/23/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-