Model Number 8888135132
Device Problem Material Integrity Problem (2978)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/15/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during use, the blue luer adapter/connector was split/cracked.Catheter was not repaired, there was no leak, saline/alcohol was the cleaning agent used.Tego was not utilized.It was stated that the device was sterilized according to specifications.There was no reported patient injury.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during use, the blue luer adapter/connector was split/cracked.Saline was used to clean the adapters.No ointments was used.No other products was being utilized with the device.Catheter was not repaired, there was no leak, saline/alcohol was the cleaning agent used.Tego was not utilized.It was stated that the device was sterilized according to specifications.Nothing unusual was observed prior to use.The catheter connector was replaced to resolve the issue.No intervention or treatment was required, no blood loss was reported.Normal dialysis was done, and procedure was completed.There was no reported patient injury.
 
Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the luer adapter was cracked, leaking, or broken.The distal end of the cannula was clamped and a water bath test was performed.Functionally, no air bubbles were detected on the red port side.It was reported that there was a leak or hole on the extension tube.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur when mishandled during clinical application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAHURKAR
Type of Device
CATHETER, HEMODIALYSIS, NON-IMPLANTED
Manufacturer (Section D)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
MDR Report Key10351329
MDR Text Key201225978
Report Number3009211636-2020-00171
Device Sequence Number1
Product Code MPB
UDI-Device Identifier10884521006218
UDI-Public10884521006218
Combination Product (y/n)N
PMA/PMN Number
K030209
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 10/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8888135132
Device Catalogue Number8888135132
Device Lot Number1921300045
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2020
Initial Date Manufacturer Received 07/15/2020
Initial Date FDA Received07/31/2020
Supplement Dates Manufacturer Received07/23/2020
09/23/2020
Supplement Dates FDA Received08/10/2020
10/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age66 YR
Patient Weight49
-
-