According to the reporter, the patient underwent a procedure for a small bowel bleed using a video capsule.The following day, it was noted that the capsule was in the patient's stomach.The capsule was retained in the patient's stomach for 12 hours.The capsule was retrieved by endoscopy.
|
Additional information: a2 a3, b1, b2, b3, b5, b6, b7, g4, h1, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The involved device was not returned, however a video was provided.Visual inspection noted the capsule functioned normally with good visualization of the segments of the gastrointestinal (gi) tract that the device did reach in the small bowel.The capsule was retained in the stomach; there was debris in the stomach.It was reported that the capsule remained in the patient for 12 hours.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: to verify passage of the capsule from the gi tract, an abdominal x-ray may be obtained at the discretion of the physician.The capsule can be removed using medical, endoscopic or surgical intervention.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|