Onsite testing by clinical engineering confirmed that the involved devices performed to specifications.The patient's historical waveforms and trend information were reviewed by clinical engineering.Findings showed vtach alarms generated for the reported episode.Clinical engineering tested all devices, tested all alarm tones, verified the monitor and passed all test.The audio settings at the central monitor volume were suspected to be low and therefore for all audio alarms were increased.There is no evidence of malfunction.This report is considered final and the issue is closed.
|
Spacelabs received a report on (b)(6) 2020 that on the (b)(6) 2020, an rn left floor and relinquished watch of patient to another rn, patient went into vtach for 20m before original rn returned.Interim rn stated she never heard the alarm.There was no reported injury to the patient in association with this incident.
|