MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX LUBRI-SIL ALL-SILICONE FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX LUBRI-SIL ALL-SILICONE FOLEY CATHETER Back to Search Results
Catalog Number 175816
Device Problems Material Puncture/Hole (1504); Device Fell (4014)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event Description
It was reported that the catheter fell out of the patient and was re-catheterised 3 times within last 5 days.The nurse re-filled one of the catheters with water that had fallen out.It was also stated that the water came out of balloon due to the hole.
 
Event Description
It was reported that the catheter fell out of the patient and was re-catheterised 3 times within last 5 days.The nurse re-filled one of the catheters with water that had fallen out.It was also stated that the water came out of balloon due to the hole.
 
Manufacturer Narrative
The reported event was inconclusive.No sample was returned for evaluation.A potential root cause for this failure could be, "tooling damaged (core pins)".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "warning: on catheter, do not use ointments or lubricants having a petroleum base.They will damage the catheter and may cause balloon to burst.After use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable laws, regulations, and policies.Keep away from sunlight.Do not store at freezing temperatures, keep at room temperature away from direct exposure to light, preferably in original box.Contains or presence of phthalates: di (2-ethylhexyl) phthalate (dehp) is a plasticizer used in some polyvinyl chloride medical devices.Dehp has been shown to produce a range of adverse effects in experimental animals, notably liver toxicity and testicular atrophy.Although the toxic and carcinogenic effects of dehp have been well established in experimental animals, the ability of this compound to produce adverse effects in humans is controversial.There is no evidence that neonates, infants, pregnant and breast feeding women exposed to dehp experience any related adverse effects.However, a lack of evidence of causation between dehp-pvc and any disease or adverse effect does not mean that there are no risks.Do not use if package is damaged.For urological use only.This is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX LUBRI-SIL ALL-SILICONE FOLEY CATHETER
Type of Device
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key10355581
MDR Text Key202541587
Report Number1018233-2020-04943
Device Sequence Number1
Product Code EZL
Combination Product (y/n)N
PMA/PMN Number
K040504
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 10/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/31/2022
Device Catalogue Number175816
Device Lot NumberNGDX3950
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/10/2020
Initial Date FDA Received08/03/2020
Supplement Dates Manufacturer Received10/07/2020
Supplement Dates FDA Received10/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-