MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ANTHOLOGY INSERTER ANT SOFT; PRSTHSS,HP,SM-CNSTRND, MTL/CRMC/PLYMR,CEMENTEDORNON-POROUS,UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ANTHOLOGY INSERTER ANT SOFT; PRSTHSS,HP,SM-CNSTRND, MTL/CRMC/PLYMR,CEMENTEDORNON-POROUS,UNCEMENTED Back to Search Results
Model Number 71365721
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/13/2020
Event Type  malfunction  
Event Description
It was reported that during efip inspection the device was noticed with a failure.Threaded tip was broken.No case involved.
 
Manufacturer Narrative
Results of investigation: the device, intended use in treatment, was returned for evaluation.A visual inspection confirmed the threaded tip of the anthology inserter ant soft is broken off and has not been returned.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHOLOGY INSERTER ANT SOFT
Type of Device
PRSTHSS,HP,SM-CNSTRND, MTL/CRMC/PLYMR,CEMENTEDORNON-POROUS,UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10356615
MDR Text Key201377177
Report Number1020279-2020-03641
Device Sequence Number1
Product Code LZO
UDI-Device Identifier03596010553409
UDI-Public03596010553409
Combination Product (y/n)N
PMA/PMN Number
K113789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 11/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71365721
Device Catalogue Number71365721
Device Lot Number11EM05624
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2020
Initial Date Manufacturer Received 07/13/2020
Initial Date FDA Received08/03/2020
Supplement Dates Manufacturer Received11/02/2020
Supplement Dates FDA Received11/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-